Description
The LED Surtron Touch 200 is an High Frequency electrosurgical equipment which is suitable for precision monopolar and bipolar surgery and micro-surgery without tissue alterations.
The unit through its performances allows pure CUT, cut-coagulation BLEND, incision with reduced production of eschar ENHANCED, superficial coagulation FORCED COAG, deep coagulation in absence of necrosis SOFT COAG, bipolar CUT and COAG. The digital reading of the delivered power and the overseeing through microcontroller of the operational functions, assure the absolute reliability of the conditions of work.
The connection of neutral electrode is constantly monitored. Safety control of patient/plate contact using split neutral electrode. The possibility to control by the handle the output functions as well as the delivery of output power, allows to implement the surgical operation without turning away the surgeon attention from the surgical field.
UNDERWATER CUTTING AND COAGULATION
Characteristics:
- Maximum power: PURE CUT 200 W – 300 Ω;
- Maximum power: BLEND CUT 150 W – 300 Ω;
- Maximum power: ENHANCED CUTTING 150 W – 300 Ω;;
- Maximum power: BIPOLAR TUR 200 W – 50 Ω;
- Maximum power: BIPOLAR PURE 120 W – 50 Ω;
- Maximum power: FORCED COAG 150 W – 200 Ω;
- Maximum power: SOFT COAG 100 W – 200 Ω;
- Maximum power: FULGURATION COAG 100 W – 1000 Ω;
- Maximum power: COAG BIPOLAR 120 W – 50 Ω;
- Maximum sealing power BIPOLAR 200 W – 50 Ω;
- Working frequency: 360 kHz;
- RF power emission time: 10 ms-30 sec.
Controls:
- OC patient board/circuit control;
- Output power control;
- Own diagnosis.